Sample preparation guideline
SRXTM at XTM beamline is based on microtomography geometry with an available x-ray energy range of 5 – 15 keV. To obtain the best imaging quality, the sample must be fixed or stabilized to make sure that there is no displacement during tomographic imaging. We highly recommend the samples be prepared in a cylindrical shape to minimize the imaging artifacts. For inorganic samples (composed of Ca, Cu, Se, Fe, Zn, etc.) or higly absorbe x-rays, the size should be less than 3 mm in diameter. However, for other samples that are composed of organic elements (C, H, O, N) or poorly absorb x-rays, the size can be up to 10 mm in diameter. SRXTM at XTM beamline is not suitable for larger samples, alloys, or titanium composites due to limited x-ray energy. Feasible samples are listed below unless stated otherwise.
SIZE of the sample
For an optimal absorption contrast, the x-ray images are aimed to obtain at 30% transmission (or 70% absorption). The absorption of X-rays by a sample depends on the quality of the X-ray beam, the elemental composition in the sample as well as the density and thickness of this sample. In general, the sample should be aligned within the FOV during rotation. This mean that the appropriate sample must be smaller than the FOV listed in the optical performance table. For example, to obtain the highest-resolution images with the pixel size of 0.72 micron, the diameter or thickness of the sample must not exceed 1.7 mm. Unless it contains mostly organic compounds and/or poorly absorbs x-rays, the sample can be larger than FOV.
In case of sample larger than FOV, the tomography dataset will be acquired from concatenated scans following the image stitching procedure. The experiment should be setup carefully to avoid misalignment that would affect the tomographic calculation. Also, the increased data volume should be taken into account.
For further analysis such as % porosity or pore distribution, we recommend that the sample should be >10-times larger than the selected feature to justify the analysis. For example, to determine the porosity, a bone sample with an average pore size of 100 microns should be prepared in 2 - 3 mm in diameter.
SHAPE of the sample
Cylindrical shaped sample is the most preferrable sample to x-ray tomography. According to the experimental setup, the x-ray beam is perpendicular to the rotation axis, the cylindrical shaped sample will give the constant pathlength (the width) during the sample rotation. This satisfies the back projection algorithm in tomographic reconstruction. In case of laminate sample, the pathlength will vary during rotation. Particularly when it turns the maximum width to the x-ray beam, it could give 0% transmission that disturbs the back projection algorithm. If it is possible, we suggest preparing the sample into a stick as illustrated.
Laminate sample should be cut to a stick.
List of feasible samples
(size in thickness or diameter)
Bones, scaffolds
-
Spongy bone < 5 mm
-
Compact bone < 3 mm
Fiber, Hair, Fabrics
Cellulose
Silk
Hair
Tooth
Less than 1 mm
Fruit seed, rice, maize
Rice 3 mm
Seed 3 - 5 mm
Wood, stem, pollen
Wood 3 - 5 mm
Stem 3 - 5 mm
Rock, Gem
Limestone/Shale 1 - 2 mm
Microfossils < 3 mm
Gem stone < 3 mm
Pearl 1 mm
Insects
-
Flies (fixed in EtOH, air dried)
-
Worm (fixed in EtOH, air dried)
Polymers, Rubbers
PLA 3 -5 mm
PP 3 - 5 mm
Composites; PVDF 3 - 5 mm
Geopolymer < 3 mm
Ceramic < 2 mm
Soils, clays
Soils
Clays < 3 mm
Available sample holders & tools
Experiment
Sample Preparation
Fixation
-
Using formaldehyde (FA) or glutaraldehyde (GA)
-
Typical 4% is used, for 12 hours or more
Dehydration & immerse
-
70% EtOH, 2 hours
-
80% EtOH, 2 hours
-
90% EtOH, 2 hours
-
100% EtOH, 2 hours
-
Immerse inHexamethyldisilazane (HMDS), 2 hours
Air dry
Make or become dry through contact with unheated air minimum 2 hours
Handling & Mounting
-
Available specimen carrier and Goniometer head
-
Sample holder options at BL1.2W
-
45° Mount, Ø = 6 mm, aluminum
-
45° and 90° Mount, Ø = 6 mm, aluminum
-
90° Mount , Ø = 6 mm , aluminum
-
Ø = 6 mm , aluminum
-
Ø = 1.5 mm, brass
-
-
Handling and mounting use container density less than sample density. Often used are small polystyrene tubes
-
In order to mount the sample to sample holder use transparent glue, wax or 2-side tape
-
User should make sure that the sample is stabilized to the sample holder other wise sample displacement during data collection will affect the CT reconstruction
Other apparatus that may be needed in sample preparation e.g. heat gun, glue gun, and stereo microscope, are also provided at the Sample Station